首页 | 本学科首页   官方微博 | 高级检索  
     


Preventive doping control analysis: liquid and gas chromatography time-of-flight mass spectrometry for detection of designer steroids
Authors:Georgakopoulos Costas G  Vonaparti Ariadni  Stamou Marianna  Kiousi Polyxeni  Lyris Emmanouil  Angelis Yiannis S  Tsoupras George  Wuest Bernhard  Nielen Michel W F  Panderi Irene  Koupparis Michalis
Affiliation:Doping Control Laboratory of Athens, Olympic Athletic Center of Athens Spiros Louis, 37 Kifissias Ave., 151 23 Maroussi, Greece. oaka@ath.forthnet.gr
Abstract:A new combined doping control screening method for the analysis of anabolic steroids in human urine using liquid chromatography/electrospray ionization orthogonal acceleration time-of-flight mass spectrometry (LCoaTOFMS) and gas chromatography/electron ionization orthogonal acceleration time-of-flight mass spectrometry (GCoaTOFMS) has been developed in order to acquire accurate full scan MS data to be used to detect designer steroids. The developed method allowed the detection of representative prohibited substances, in addition to steroids, at concentrations of 10 ng/mL for anabolic agents and metabolites, 30 ng/mL for corticosteroids, 500 ng/mL for stimulants and beta-blockers, 250 ng/mL for diuretics, and 200 ng/mL for narcotics. Sample preparation was based on liquid-liquid extraction of hydrolyzed human urine, and the final extract was analyzed as trimethylsilylated derivatives in GCoaTOFMS and underivatized in LCoaTOFMS in positive ion mode. The sensitivity, mass accuracy, advantages and limitations of the developed method are presented.