首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Precision of Flash Pyrolysis-GC-MS Analysis: Experimental Evaluation of the Sources of Variability
Authors:Lei Ji  Lyadi Pio Riana  Michel Sablier  Jérôme Vial
Institution:1. Laboratoire des Mécanismes Réactionnels, Ecole Polytechnique, CNRS, 91128, Palaiseau Cedex, France
2. Laboratoire Environnement et Chimie Analytique, ESPCI, CNRS, 10, rue Vauquelin, 75005, Paris, France
Abstract:The present study was carried out to evaluate the precision and to identify significant sources of variability in flash pyrolysis-GC-MS experiments. The analysis of variance has been applied to the example of polyethylene pyrolysis to estimate the impact of several controlled factors: sample handling, flash pyrolysis filament ageing and calibration. It was demonstrated that to get the best precision in experimental results, it is recommended to set the operator to reduce the impact of sample handling procedure and the filament to avoid variability due to changes in filament characteristics. In comparison, filament ageing and filament calibration are shown to influence in a limited extent the experimental results.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号