首页 | 本学科首页   官方微博 | 高级检索  
     检索      


PEG-modified glucose oxidase immobilized on a PVA cryogel membrane for amperometric biosensor applications
Authors:Doretti L  Ferrara D  Gattolin P  Lora S  Schiavon F  Veronese F M
Institution:Ufficio Sicurezza e Prevenzione del CNR, Corso Stati Uniti 4, 35127 Padova, Italy.
Abstract:Poly(ethylene glycol)-modified glucose oxidase was immobilized in a poly(vinyl alcohol) cryogel membrane, obtained by a freezing-thawing cyclic process, to obtain a suitable amperometric glucose sensor. The covalent linkage between PEG and GOD molecule improved the physical immobilization of enzyme in the polymeric matrix, by decreasing its loss in time. Sensor behaviour was evaluated electrochemically with a hydrogen peroxide electrode. The glucose content in standard solutions was determined and linear calibration curves in the 5x10(-5)-5x10(-3) mol l(-1) range were obtained. The kinetic parameters in the immobilized system were evaluated and analytical characteristics of sensor, including stability and influence of pH and temperature, were determined.
Keywords:
本文献已被 PubMed 等数据库收录!