MRI measurements of the dependence on T1 of the echo amplitudes using a multiple spin-echo scheme
Authors:A Fransson  A Ericsson  GO Sperber
Institution:

?Department of Diagnostic Radiology, Uppsala University, Uppsala, Sweden

?Department of Physiology and Medical Biophysics, Uppsala University, Uppsala, Sweden

*Department of Radiation Physics, Karolinska Institutet, Stockholm, Sweden

aStockholm University, Stockholm, Sweden

bDepartment of Hospital Physics, University Hospital, Uppsala, Sweden

Abstract:Analytical calculations using the Bloch formalism were performed to assess the dependence on T1 of the echo amplitudes for the Phase-Alternating Phase-Shift (PHAPS) multiple spin-echo protocol. Measurements in a 0.5 T MR imaging unit were performed to ratify the analytical results. especially for low T2 values, the echo amplitudes were erroneous, with an increasing contribution from stimulated echo components with increasing T1. Apart from affecting T2 estimates, stimulated echoes generated a non-monoexponential signal decay of the echo trains. The results confirmed previous simulation studies as regards the dependence on T1 of T2 estimates from PHAPS.
Keywords:MRI  Relaxation time measurements  Multiple echo  Stimulated echoes