Development and Validation of a Liquid Chromatographic Method for Determination of Efavirenz in Human Plasma |
| |
Authors: | A. Lakshmi Sailaja K. Kishore Kumar D. V. R. Ravi Kumar C. Mohan Kumar N. M. Yugandhar G. Srinubabu |
| |
Affiliation: | (1) Centre for Biotechnology, Andhra University College of Engineering, Visakhapatnam, 530003, India |
| |
Abstract: | A simple, rapid, and sensitive high-performance liquid chromatographic method for estimation of efavirenz in human plasma has been developed and validated. Chromatography was performed with C18 analytical column and 50:50 acetonitrile–phosphate buffer (pH 3.5) as mobile phase. Compounds were monitored by UV detection at 247 nm. The retention time for efavirenz was 6.45 min and that for the internal standard, nelfinavir, was 2.042 min. Response was a linear over the concentration range of 0.1 μg–10 μg mL−1 in human plasma. The method was simple, specific, precise and accurate and was useful for bioequivalence and pharmacokinetic studies of efavirenz. |
| |
Keywords: | Column liquid chromatography Validation Human plasma Efavirenz |
本文献已被 SpringerLink 等数据库收录! |
|