Sensitive Determination of 2‐Methoxyestradiol in Pharmaceutical Preparations and Biological Fluids by KMnO4‐Na2SO3 Chemiluminescence System |
| |
Authors: | Hanchun Yao Jie Zhou Min Zhang Erjun Xu Juan Chen Zhenzhong Zhang |
| |
Affiliation: | School of Pharmaceutical Sciences, Zhengzhou University, Zhengzhou 450001, Henan, P. R. China |
| |
Abstract: | A simple and novel flow‐injection chemiluminescence (FI‐CL) method was established for the determination of 2‐Methoxyestradiol (2‐ME) in pharmaceutical preparations and biological fluids. The method was based on the significant enhancement of the CL from the KMnO4‐Na2SO3 reaction by 2‐ME in acidic medium. Under optimized conditions, the CL intensity was correlated linearly with concentration of 2‐ME in the range of 5.0 × 10?8‐5.0 × 10?6 M (r = 0.9995). The detection limit (3σ) of 2‐ME was 7.5 × 10?9 M and the relative standard deviation was 0.8% at 5.0 × 10?7 M 2‐ME (n = 8). The proposed method was successfully applied for the flow‐injection CL determination of 2‐ME in pharmaceutical preparations and biological fluids with the recoveries from 92.4 to 106.8%. The possible CL reaction mechanism was also discussed briefly. |
| |
Keywords: | Biological fluids Flow‐injection chemiluminescence Potassium permanganate Sodium sulfite 2‐Methoxyestradiol |
|