首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Voltammetric Determination of Boric Acid Using the Ternary Complexation System with Salicylaldehyde and H‐Acid
Authors:Kô Takehara  Takao Fujimori  Kei Inagi  Mari Kajiwara  Yoshikazu Harata  Kazuhisa Yoshimura
Institution:1. Department of Chemistry, Faculty of Sciences, Kyushu University, 6‐10‐1, Hakozaki, Fukuoka 812‐8581, Japan;2. Department of Chemical Science and Engineering, Miyakonojo National College of Technology, 473‐1, Yoshio‐cho, Miyakonojo, Miyazaki 885‐8567, Japan;3. Takaoka Chemical Co., ltd., 80 Nagare, Jimokuji, Ama‐city, Aichi 490‐1111, Japan
Abstract:A square‐wave voltammetric method for the determination of boric acid in water has been described based on the new understanding of the electrochemical behavior of boric acid‐Azomethine H complexation. Salicylaldehyde and H‐acid were used as the starting materials of boric acid‐Azomethine H complex and their concentrations were optimized for boric acid determination in water. A glassy carbon electrode, instead of a conventional mercury electrode, was used in the measurement. The detection limit of the proposed method was 0.10 mg B dm?3. The proposed method was successfully used for boric acid determination in the water from a seawater desalination RO plant.
Keywords:Boron determination  Boric acid‐Azomethine H complex  Square‐wave voltammetry  RO‐desalinated water
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号