首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and Validation of an Stability Indicating RP-LC Method for Determination of Imatinib Mesylate
Authors:Girish Bende  Sivacharan Kollipara  Venugopal Kolachina  Ranendra Saha
Institution:1. Formulation Development and Pharmacokinetics Laboratory, Pharmacy Group, Birla Institute of Technology and Science, Pilani, 333031, Rajasthan, India
Abstract:A new, sensitive and stability indicating liquid chromatographic method has been developed for the determination of imatinib mesylate (IM). Efficient chromatographic separation was achieved using a C18 column with simple mobile phase combination delivered in an isocratic mode and quantitation was carried out using ultraviolet detection. For the first time, a novel microwave assisted degradation procedure was employed for stress testing studies. In addition, orthogonal separation technique was applied to demonstrate selectivity of the proposed method. The method has demonstrated excellent linearity over the range of 25–1,600 ng mL?1. Moreover, the method was found to be sensitive with a low limit of detection (3.35 ng mL?1) and limit of quantitation (10.16 ng mL?1). The method has shown good and consistent recoveries (99.35–100.69%) with low intra- and inter-day relative standard deviation (RSD) (<2.5%). Experimental design confirmed that peak area was unaffected by small changes in critical factors, in robustness study. The validated method was successfully applied for determination of IM in pharmaceutical formulations.
Keywords:
本文献已被 SpringerLink 等数据库收录!