首页 | 本学科首页   官方微博 | 高级检索  
     


Performance and testing of a four channel high-resolution heterodyne interferometer
Authors:J.J.J. Dirckx  H.J. van Elburg  W.F. Decraemer  J.A.N. Buytaert  J.A. Melkebeek
Affiliation:1. Laboratory of Biomedical Physics, University of Antwerp, Groenenborgerlaan 171, 2020 Antwerp, Belgium;2. EESA-EELAB, Ghent University, Sint-Pietersleeuwstraat 41, 9000 Ghent, Belgium
Abstract:When studying complex vibrations, simultaneous measurements at several points are indispensable if one is dealing with objects whose vibrational behavior is not guaranteed to be stable over longer periods of time, such as biological specimens, micro-mechanical elements or objects characterized by nearly resonant normal modes with different vibrational patterns. Obviously, both amplitude and phase need to be measured at each point to obtain a full characterization of the vibration. We introduce birefringent beam displacers as a highly efficient beam multiplying method to create a system of four heterodyne interferometers operating in parallel from a single laser source. The design and the performance characteristics (resolution, cross-talk) of the instrument will be discussed. The system revealed the existence of running vibrational modes on an electrically driven plate clamped along its outer edge.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号