Affiliation: | aMedical Physicochemical Centre, Donetsk Medical University, 16 Ilych Avenue, Donetsk 83003, Ukraine bSINTERFACE Technologies, Volmerstraße 7, 12489 Berlin, Germany cMax-Planck-Institut für Kolloid- und Grenzflächenforschung, Am Mühlenberg 1, 14476 Potsdam/Golm, Germany |
Abstract: | One of the essential differences in the design of bubble pressure tensiometers consists in the geometry of the measuring capillaries. To reach extremely short adsorption times of milliseconds and below, the so-called deadtime of the capillaries must be of the order of some 10 ms. In particular, for concentrated surfactant solutions, such as micellar solutions, short deadtimes are needed to minimize the initial surfactant load of the generated bubbles. A theoretical model is derived and confirmed by experiments performed for a wide range of experimental conditions, mainly in respect to variations in deadtime and bubble volume. |