首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of quantitative performance of three fluorescence labels in CE/LIF analysis of aspartate and glutamate in brain microdialysate
Authors:Wagner Zsolt  Tábi Tamás  Zachar Gergely  Csillag András  Szöko Eva
Affiliation:Department of Pharmacodynamics, Semmelweis University, Budapest, Hungary.
Abstract:
Three different fluorescent tags have been compared for the quantitative analysis of aspartate and glutamate in brain microdialysate samples. Separation conditions have been optimized to achieve short analysis time using reversed polarity separation in coated capillary. Method validation has revealed similar quantification limit of 0.1?μM of analytes using either of the labels, although LOD values were different: 7.8-9.8?nM for 4-fluoro-7-nitro-2,1,3-benzoxadiazole, 3.5?nM for fluorescein-5-isothiocyanate and 1.3-1.5?nM for carboxyfluorescein succinimidyl ester derivatives. The almost two orders of magnitude difference between LOD and LOQ values is likely due to the unreliable derivatization reaction at low sample concentration. Based on the superior stability, FITC derivatization was used for the analysis of biological samples. The applicability of the method has been demonstrated by analyzing basal and potassium evoked amino acid concentrations in individual brain microdialysate samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏