首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Improvement in performance of a flow electrochemical sensor by using carbamoyl-arms polyazamacrocycle for the preconcentration of lead ions onto the electrode
Authors:Rihab Nasraoui  Didier Floner  Florence Geneste
Institution:Université de Rennes 1, UMR-CNRS 6226, Laboratoire des Sciences Chimiques de Rennes, Equipe catalyse et organométalliques, Campus de Beaulieu, 35042 Rennes cedex, France
Abstract:A flow electrochemical sensor for trace analysis of lead, using TETRAM-modified graphite felt electrode is reported here. TETRAM ligands are covalently immobilized on the graphite felt by chemical reactions on amino acid linkers, previously attached to the electrode by an electrochemical process. The detection is performed in two steps: the preconcentration of Pb2+ ions by complexation with immobilized TETRAM and the analysis by linear sweep stripping voltammetry. A calibration curve typical of at least two equilibrium processes is obtained. A limit of detection of 2.5 × 10?8 mol L?1 is reached for a total analysis time of 35 min. Interestingly, the flow sensor shows a good selectivity toward lead in presence of Cu2+, Cd2+, Ni2+, Zn2+ and Co2+ ions. This new sensor exhibits improved sensitivity and selectivity compared to the previously reported sensor using cyclam-modified electrode. It is stable after three uses, using strong acidic medium for the regeneration step.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号