Analysis of amides in the air by sampling with SPME and detection by GC-FID |
| |
Authors: | Parreira Fabrício Vilela de Souza Bergo Patrícia Luísa Rabelo de Carvalho Ciomara de Lourdes Cardeal Zenilda |
| |
Affiliation: | Chemistry Department, ICEx, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil. |
| |
Abstract: | This paper proposes a solid-phase microextraction (SPME) method for the passive sampling and determination of N,N-dimethylacetamide (DMA) and N,N-dimethylformamide (DMF) in air by capillary GC with flame ionization detection. The optimized conditions for SPME method were grab sampling, polydimethylsiloxane/divinylbenzene fiber, extraction for 90 min at 25 +/- 2 degrees C, desorption for 3 min at 220 degrees C, and relative humidity 45 +/- 2%. Under these conditions, the method presented good linearity (R = 0.996), repeatability (%RSD 2.79 and 9.85 for DMF and DMA, respectively), and detection limit (0.021 and 0.024 mg/m3 for DMF and DMA, respectively). |
| |
Keywords: | N,N‐Dimethylacetamide N,N‐Dimethylformamide SPME passive sampling GC‐FID |
本文献已被 PubMed 等数据库收录! |
|