Indirectly illuminated X-ray area detector for X-ray photon correlation spectroscopy |
| |
Authors: | Shinohara Yuya Imai Ryo Kishimoto Hiroyuki Yagi Naoto Amemiya Yoshiyuki |
| |
Institution: | Graduate School of Frontier Sciences, The University of Tokyo, 5-1-5 Kashiwanoha, Kashiwa, Chiba 277-8561, Japan. yuya@k.u-tokyo.ac.jp |
| |
Abstract: | An indirectly illuminated X-ray area detector is employed for X-ray photon correlation spectroscopy (XPCS). The detector consists of a phosphor screen, an image intensifier (microchannel plate), a coupling lens and either a CCD or CMOS image sensor. By changing the gain of the image intensifier, both photon-counting and integrating measurements can be performed. Speckle patterns with a high signal-to-noise ratio can be observed in a single shot in the integrating mode, while XPCS measurement can be performed with much fewer photons in the photon-counting mode. By switching the image sensor, various combinations of frame rate, dynamic range and active area can be obtained. By virtue of these characteristics, this detector can be used for XPCS measurements of various types of samples that show slow or fast dynamics, a high or low scattering intensity, and a wide or narrow range of scattering angles. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|