首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Study on open tubular capillary affinity liquid chromatography
Authors:Ye M  Zou H  Liu Z  Lei Z  Ni J
Institution:National Chromatographic R. & A. Center, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, China.
Abstract:A novel mode of affinity chromatography (AC) based on an open tubular capillary column (OTAC) is demonstrated. The OTAC column is prepared by immobilizing Cibacron blue F3GA onto the inner surface of a 50-microm-i.d. capillary column. The AC experiment is performed on a capillary electrophoresis instrument by using its pressure system as the driving force. Bovine serum albumin and lysozyme (Lys) are successfully separated with stepwise gradient elution. The relative standard deviation (RSD) for the elution time of the retained Lys is 0.08%, and good repeatability of its peak area and peak height with an RSD value lower than 2.12% for 10 consecutive runs is observed. The loading capacity and detection limit for the retained Lys are approximately 36 ng and 8.6 ng, respectively. It is also found that the amount of protein adsorbed is unaffected by the flow rate of the loading buffer, and OTAC can be used for the fast determination of biopolymers. Some of the advantages of OTAC over conventional modes of open tubular capillary liquid chromatography are that the detection sensitivity and loading capacity of a sample can be greatly improved, because the relatively large inner diameter of the capillary can be adopted and the whole capillary column can be used to adsorb the solute in OTAC.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号