Intraocular lens design for treating high myopia based on individual eye model |
| |
Authors: | Yang Wang Zhaoqi Wang Yan Wang |
| |
Affiliation: | a Institute of Modern Optics, Nankai University, Tianjin 300071, PR China b Key Laboratory of Opto-electronic Information Science and Technology, Ministry of Education, China c Tianjin Eye Hospital, Refractive Surgery Center, Tianjin 300020, PR China |
| |
Abstract: | In this research, we firstly design the phakic intraocular lens (PIOL) based on individual eye model with optical design software ZEMAX. The individual PIOL is designed to correct the defocus and astigmatism, and then we compare the PIOL power calculated from the individual eye model with that from the experiential formula. Close values of PIOL power are obtained between the individual eye model and the formula, but the suggested method has more accuracy with more functions. The impact of PIOL decentration on human eye is evaluated, including rotation decentration, flat axis decentration, steep axis decentration and axial movement of PIOL, which is impossible with traditional method. To control the PIOL decentration errors, we give the limit values of PIOL decentration for the specific eye in this study. |
| |
Keywords: | Phakic intraocular lens Eye model Wavefront aberration Decentration ZEMAX |
本文献已被 ScienceDirect 等数据库收录! |
|