首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analysis of microgram amounts of particulate material by simultaneous multiwavelength AES
Authors:Robert S Pomeroy  Jeffrey D Kolczynski  Jonathan V Sweedler  M Bonner Denton
Institution:(1) Department of Chemistry, University of Arizona, 85721 Tucson, AZ, USA
Abstract:A unique simultaneous emission spectrograph is utilized to perform qualitative and quantitative analysis on trace quantities of solid particulates. The atomic emission spectroscopic system consists of a direct current plasma source and an echelle spectrograph with a charge injection device detector, enabling the system to simultaneously measure the wavelength range from 220 nm to 520 nm with 0.02 nm resolution at 300 nm. Monitoring all wavelengths simultaneously allows the qualitative and quantitative determination of most major and minor constituent in a trace quantity of sample with little prior knowledge about the sample. The ability to perform qualitative and quantitative analysis on particulates is demonstrated by evaluating NBS certified coal fly ash, as well as a sample taken from the respirator air filter at an acute care unit in a hospital.Presented in part at the 1989 European Winter Conference on Plasma Spectrochemistry, Reutte, Austria
Keywords:atomic emission spectroscopy  charge transfer device  CID detection  trace elemental analysis
本文献已被 SpringerLink 等数据库收录!