Accurate depth-independent determination of acoustic backscatter coefficients with focused transducers |
| |
Authors: | T J Hall E L Madsen J A Zagzebski E J Boote |
| |
Affiliation: | Department of Medical Physics, University of Wisconsin, Madison 53706. |
| |
Abstract: | ![]() The accuracy of a method of data reduction for determining acoustic backscatter coefficients was tested using focused transducers and narrow-band pulses. Two phantoms with well-defined scattering properties were the bases of the tests, one having low attenuation and one with tissue-mimicking attenuation. The experimentally determined backscatter coefficients were found to be independent of transducer-to-scattering-volume distance and to agree very well with theoretical values, typically within 10%. |
| |