首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Cloud point extraction, preconcentration and simultaneous spectrophotometric determination of nickel and cobalt in water samples
Authors:Safavi A  Abdollahi H  Hormozi Nezhad M R  Kamali R
Institution:Department of Chemistry, College of Sciences, Shiraz University, Shiraz 71454, Iran. safavi@chem.susc.ac.ir
Abstract:Cloud point extraction has been used for the preconcentration and simultaneous spectrophotometric determination of nickel and cobalt after the formation of a complex with 2-amino-cyclopentene-1-dithiocarboxylic acid (ACDA), and latter analysis by spectrophotometer using Triton X-114 as surfactant. The parameters affecting the separation phase and detection process were optimized. Under the optimum experimental conditions (i.e. pH=5, 0.07 mM ACDA, Triton X-114=0.25% (w/v)), calibration graphs were linear in the range of 20-500 and 20-200 microg l(-1) with detection limits of 10 and 7.5 microg l(-1) for Ni and Co, respectively. The method was applied to the determination of Ni and Co in natural and waste water samples with satisfactory results.
Keywords:
本文献已被 PubMed 等数据库收录!