The effect of the source pressure on the abundance of ions of noncovalent protein assemblies in an electrospray ionization orthogonal time-of-flight instrument |
| |
Authors: | Tahallah N Pinkse M Maier C S Heck A J |
| |
Affiliation: | Department of Biomolecular Mass Spectrometry, Bijvoet Center for Biomolecular Research and Utrecht Institute for Pharmaceutical Sciences, Utrecht University, Sorbonnelaan 16, 3584 CA Utrecht, The Netherlands. |
| |
Abstract: | The effect of elevating the pressure in the interface region of an electrospray ionization orthogonal time-of-flight mass spectrometer on the ion intensity of different noncovalent protein assemblies has been investigated. Elevating the pressure in the interface region generally led to an enhanced detection of high m/z ions. The optimum pressure was found to be dependent on the m/z value of the ions. This pressure effect should be carefully addressed when relating ion abundance in the mass spectra to solution phase abundance of noncovalent protein assemblies. |
| |