首页 | 本学科首页   官方微博 | 高级检索  
     


Simultaneous Determination of 11 Elements in Fly Ash by Inductively Coupled Plasma Orthogonal Acceleration Time-of-Flight Mass Spectrometry After Closed-Vessel Microwave-Assisted Extraction with Ammonium Fluoride
Authors:Lenka Husáková  Iva Urbanová  Tereza Šídová  Michaela Šafránková
Affiliation:1.Department of Analytical Chemistry, Faculty of Chemical Technology,University of Pardubice,Pardubice,Czech Republic
Abstract:
A method for simultaneous multi-element analysis of fly ash samples by inductively coupled plasma orthogonal acceleration time-of-flight mass spectrometry (oaTOF-ICP-MS) after closed-vessel microwave extraction with ammonium fluoride was introduced here. Corrosive and/or toxic acids like HF, HCl or HClO4, as well as HNO3, which are commonly used during sample preparation of the fly ash samples, are avoided in this method. The spectral effects due to the formation of different Cl, Na, K, Ca, Mg-containing polyatomic species interfering with the determination of a number of elements like As, Se or Ni during the oaTOF-ICP-MS analysis are negligible. Under the optimum experimental extraction conditions evaluated using a fractional factorial design (10 mg of the sample extracted with 5 mL of 140 g/L NH4F for 10 min at 200°C), analysis of the resulting supernatant with Rh as an internal standard enabled precise and accurate simultaneous determination of 11 elements (Li, Be, Ni, As, Se, Rb, Sb, Cs, W, Tl and U) at trace and ultratrace levels. The accuracy was assessed by analysing two certified reference materials, namely Fine Fly Ash CTA-FFA-1 and Constituent Elements in Coal Fly Ash Standard Reference Material® 1633b. The precision of the reported method was better than 10%.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号