Spectral Calibration for the Large Aperture Infrared Radiometer
YE Zhao,WAN Zhi,LI Xian-sheng,REN Jian-wei,LIU Hong-xing.Spectral Calibration for the Large Aperture Infrared Radiometer[J].Spectroscopy and Spectral Analysis,2012,32(7):1994-1998.
Authors:YE Zhao  WAN Zhi  LI Xian-sheng  REN Jian-wei  LIU Hong-xing
Institution:1. Changchun Institute of Optics,Fine Mechanics and Physics,Chinese Academy of Sciences,Changchun 130033, China2. Graduate University of Chinese Academy of Sciences,Beijing 100049,China
Abstract:The present paper introduced a spectral calibration method to calibrate a large aperture infrared radiation facility. We developed a radiometer which consisted of optical system, infrared detectors(cavity pyroelectric detector and HgCdTe detector 2~14 μm), fine mechanical modulator, lock in amplifier, signal processor, etc. At first, we analyzed the method on how to measure the spectral calibration of the large aperture infrared radiometer, and established the spectral calibration facility. Then, we tested the nonlinear response for the cavity pyroelectric detector and HgCdTe detector. Finally, we used the cavity pyroelectric detector to calibrate the relative spectral responsivity of HgCdTe detector at several wavelengths on the facility. Through the comparison of the two methods for measuring the relative spectral responsivity, the average of multiple measurements and comparative analysis of two methods were given. The uncertainty analysis of the whole system showed that the measurement uncertainty of the facility was better than 3.4%.
Keywords:Large aperture  Infrared radiometer  Spectral calibration  Cavity pyroelectric detector  HgCdTe detector  