Ultramicro determination of organic halogens using sealed tube dry combustion method |
| |
Authors: | Yoshihide Tanaka Ayako Okazaki Kei Hozumi |
| |
Affiliation: | (1) Kyoto Pharmaceutical University, Yamashina, 607 Kyoto, Japan |
| |
Abstract: | A sealed tube dry combustion method that is simple, accurate, and precise for ultramicro determination of halogens in organic compounds has been developed. Samples below 1 mg are heated at 580 °C for 1 h in an electric furnace. After the combustion and absorption of the gaseous halogen, the interior of the tube is rinsed. Subsequent titration is carried out potentiometrically with silver nitrate and yielded standard deviations of 0.21% for chlorine, 0.17% for bromine, and 0.25% for iodine. |
| |
Keywords: | ultramicro determination of halogens organic halogens sealed tube dry combustion potentiometric titration of halides |
本文献已被 SpringerLink 等数据库收录! |