Ultrasonic density measurement cell design and simulation of non-ideal effects
Authors:Higuti Ricardo Tokio  Buiochi Flávio  Adamowski Júlio Cezar  de Espinosa Francisco Montero
Institution:UNESP - Universidade Estadual Paulista, Departamento de Engenharia Elétrica, Avenida Brasil 56, 15385-000 Ilha Solteira, SP, Brazil. tokio@dee.feis.unesp.br
Abstract:This paper presents a theoretical analysis of a density measurement cell using an unidimensional model composed by acoustic and electroacoustic transmission lines in order to simulate non-ideal effects. The model is implemented using matrix operations, and is used to design the cell considering its geometry, materials used in sensor assembly, range of liquid sample properties and signal analysis techniques. The sensor performance in non-ideal conditions is studied, considering the thicknesses of adhesive and metallization layers, and the effect of residue of liquid sample which can impregnate on the sample chamber surfaces. These layers are taken into account in the model, and their effects are compensated to reduce the error on density measurement. The results show the contribution of residue layer thickness to density error and its behavior when two signal analysis methods are used.
Keywords:43  35  Yb  43  35  Zc  43  58  Bh