首页 | 本学科首页   官方微博 | 高级检索  
     


A new quality control method for lateral flow assay
Authors:Li Huang  Di Zhang  Lirong Jiao  Enben Su  Nongyue He
Affiliation:State Key Laboratory of Bioelectronics, School of Biological Science and Medical Engineering, Southeast University, Nanjing 210096, China; National Demonstration Center for Experimental Biomedical Engineering Education, Southeast University, Nanjing 210096, China; Getein Biotechnology Co., Ltd., Nanjing 210000, China
Abstract:
Precision and repeatability are challenging issues in point of care testing (POCT) analysis. Herein, we proposed a lateral flow assay (LFA) based on internal quality control microspheres to realize the accurate diagnosis of HbA1c in human body. Fluorescein cy5 decorated microspheres are used as labels for HbA1c detection, and BSA- fluorescein isothiocyanate (FITC) decorated microspheres are used as internal quality control labels. One test line was employed in the strip for the detection of glycosylated hemoglobin (HbA1c). This method can eliminate the interference of environmental factors (temperature, humidity, etc.) to LFA in the process of chromatography, and improve the precision and accuracy of HbA1c detection. The CV for detection of low concentration HbA1c was 1.05%, and the CV for detection of high concentration HbA1c was 0.69%. We envision the method to have great prospect in in vitro diagnosis (IVD).
Keywords:Point of care testing  Lateral flow assay  HbA1c  Fluorescein isothiocyanate  Precision
本文献已被 ScienceDirect 等数据库收录!
点击此处可从《中国化学快报》浏览原始摘要信息
点击此处可从《中国化学快报》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏