首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Strontium(II) selective PVC membrane electrode based acetophenone semicarbazone (ACS) as an ionophore
Institution:1. Department of Chemistry, Zakir Husain College (University of Delhi) J.L. Nehru Marg, New Delhi 110 002, India;2. Department of Applied Chemistry, MAIT, (G G S I P University) Sec-22 Rohini, New Delhi, India
Abstract:A new PVC membrane based strontium(II) ion-selective electrode has been constructed using acetophenone semicarbazone as a neutral carrier. The sensor exhibits a Nerstian response for strontium(II) ion over a wide concentration range 1.0 × 10?2–1.0 × 10?7 M with the slope of 29.4 mV/per decade. The limit of detection was 2.7 × 10?8 M. It was relatively fast response time (<10 s for concentration ?1.0 × 10?3 and <15 s for concentration of ?1.0 × 10?6 M) and can be used for 8 months without any considerable divergence in potentials. The proposed sensor revealed relatively good selectivity and high sensitivity for strontium(II) over a mono, di, trivalent cation and can be used in a pH range of 2.5–10.5. It was also successfully used as an indicator electrode in potentiometer titration and in the analysis of concentration in various real samples.
Keywords:Strontium  Strontium(II) ion-electrode  PVC membrane  Ionophore  Strontium nitrate  Semicarbazidehydrochloride  Potentiometry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号