Comments on the primary and scatter dose-spread kernels used for convolution methods |
| |
Authors: | Akira Iwasaki |
| |
Affiliation: | School of Health Sciences, Hirosaki University, 66-1 Hon-cho, Hirosaki, Aomori 036-8564, Japan |
| |
Abstract: | The basic primary and scatter dose-spread kernels used for convolution methods are usually produced by Monte Carlo simulations with the interaction point forced to the center of a large water phantom. However, it is still not clear whether such Monte Carlo based kernels allow accurate dose calculations with a wide range of field sizes and depths, especially in thorax phantoms. Using the differential primary and scatter concept, this paper proposes another type of basic kernel, with which perfectly accurate primary and scatter absorbed dose calculations can be performed under conditions that the beam is parallel, the incident beam intensity is uniform within and zero outside the field, and the primary beam attenuation coefficient along raylines is not a function of depth and off-axis distance. |
| |
Keywords: | Convolution method Primary dose-spread kernel Scatter dose-spread kernel |
本文献已被 ScienceDirect 等数据库收录! |
|