D-dot monitor for sub-nanosecond high voltage pulse measurement
Wei Bing , Fang Dongfan , Qing Yanling , Sun Qizhi , Feng Shuping , Yang Libing. D-dot monitor for sub-nanosecond high voltage pulse measurement[J]. High Power Laser and Particle Beams, 2012, 24(6): 1497-1501. DOI: 10.3788/HPLPB20122406.1497
Authors:Wei Bing    Fang Dongfan    Qing Yanling    Sun Qizhi    Feng Shuping    Yang Libing
Affiliation:1.Institute of Fluid Physics,CAEP,P.O.Box 919-108,Mianyang 621900,China
Abstract:A kind of D-dot monitor has been designed to measure the high voltage pulse on an impedance transformer line of compact fast rise time pulse generator. The scaling factor of the D-dot monitor was calibrated in-situ by using a calibrated resistive divider and a high voltage pulse whose rise time is about 50 ns. The response of the monitor was tested by using a 50 Ω transmission line installed with the D-dot monitor. When the rise time of input pulse is about 320 ps, the rise time of the pulse output by the D-dot monitor is about 350 ps. The response of the D-dot monitor is about 150 ps. Thus the monitor can measure the voltage pulses with the rise time exceeding 450 ps. The experiment results indicate that the D-dot monitor can reliably and accurately measure the sub-nanosecond high voltage pulses.
Keywords:sub-nanosecond high voltage pulse  D-dot monitor  frequency response  calibration