Abstract:Spectrophotometric multicomponent analysis is considerd on the basis of inverse multivariate calibration with linear methods (ordinary least squares, principal component, ridge and partial least squares regression) and with the non-linear methods ACE and the non-linear partial least squares. The performance of the different methods is compared by paired F-tests. As an estimate of the error variance the residual mean sum of squares in the analysis of variance table is used. The comparison is demonstrated for the infrared spectrometric analysis of the hydroxyl group content of brown coal measured in diffuse reflectance. Although the error variances among the calibration methods differ gradually, the differences are much less pronounced at statistical level.