首页 | 本学科首页   官方微博 | 高级检索  
     检索      


New EDTA determination method based on ion chromatography with suppressed conductimetric detection
Authors:György Pátzay  Yahya Ramadan  Csilla Tonkó
Institution:1. Department of Chemical and Environmental Process Engineering, Budapest University of Technology and Economics, Budafoki út 8, Budapest, 1111, Hungary
Abstract:A novel direct method for the determination of EDTA in alkaline radioactive evaporator residue solution was developed and validated based on ion chromatography with suppressed conductimetric detection and anion exchange columns (A Supp 4, 4 mm × 250 mm and A Supp 5, 4 mm × 150 mm). The yttrium-EDTA complex resulted one single chromatographic peak in the eluent and allowed the correct determination of EDTA in an alkaline, high concentration radioactive waste water. Depending on coexisting substances, suitable eluent is 10.0 mM carbonate buffer/pH 10.6 or 10.75 (t R,Y–EDTA = 7.01 and 6.4 min, respectively). For 10.0 mM carbonate buffer/pH 10.6 and isocratic flow rate of 1.0 cm3/min, a linear calibration curve was obtained from 5 to 40 mg/dm3 (r > 0.999) EDTA. Good resolution was achieved from commonly coexisting anions (chloride, nitrite, nitrate, sulphate, phosphate, bromide and citrate). The developed simple ion chromatographic method was applied for the assay of EDTA in various radioactive alkaline solutions.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏