Comparison of experimental and calculated peak shapes for three cylindrical geometry FAIMS prototypes of differing electrode diameters
Authors:Email author" target="_blank">Roger?GuevremontEmail author  Randy?Purves
Institution:Ionalytics Corporation, Ottawa, Ontario K1A 0R6, Canada. roger.guevremont@ionalytics.com
Abstract:High-field asymmetric waveform ion mobility spectrometry (FAIMS) separates ions at atmospheric pressure and room temperature based on the difference of the mobility of ions in strong electric fields and weak electric fields. This field-dependent mobility of an ion is reflected in the compensation voltage (CV) at which the ion is transmitted through FAIMS, at a given asymmetric waveform dispersion voltage (DV). Experimental CV, relative peak ion intensity, and peak width data were compared for three FAIMS prototypes with concentric cylindrical electrodes having inner/outer electrode radii of: (1) 0.4/0.6 cm, (2) 0.8/1.0 cm, and (3) 1.2/1.4 cm. The annular analyzer space was 0.2 cm wide in each case. A finite-difference numerical computation method is described for evaluation of peak shapes and widths in a CV spectrum collected using cylindrical geometry FAIMS devices. Simulation of the radial distribution of the ion density in the FAIMS analyzer is based upon calculation of diffusion, electric fields, and the electric fields introduced by coulombic ion-ion repulsion. Excellent agreement between experimental and calculated peak shapes were obtained for electrodes of wide diameter and for ions transmitted at low CV.