首页 | 本学科首页   官方微博 | 高级检索  
     


Improved short-lived nuclide activation analysis
Authors:N. N. Papadopoulos  N. M. Spyrou  N. F. Tsagas  G. E. Hatzakis
Affiliation:(1) National Centre for Scientific Research "ldquo"Demokritos"rdquo", 153 10 Aghia Paraskevi, Athens, (Greece);(2) Department of Physics, University of Surrey, GU2 5XH Guildford, Surrey, (UK);(3) Democritus University of Thrace, 67100 Xanthi, (Greece)
Abstract:
A novel technique has been developed to increase the sensitivity, precision and accuracy of short-lived nuclide activation analysis and especially of delayed fission neutron counting, by improving the counting statistics. This has been achieved by approaching the irradiated sample to the detector during the counting period in order to compensate for the rapid radioactive decay by variation of the counting efficiency, permitting prolongation of the counting period with constant count rate. The sample holder movement with approximately exponentially decreasing velocity is implemented by a variable DC motor with a rotary to linear movement converter, which is being replaced by a more reliable stepping motor.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号