Highly-sensitive NO,NO_2,and NH_3 measurements with an open-multipass cell based on mid-infrared wavelength modulation spectroscopy |
| |
Abstract: | A compact prototype based on mid-infrared wavelength modulation spectroscopy(WMS)is developed for the simul-taneous monitoring of NO,NO_2,and NH_3 in the urban area.Three quantum cascade lasers(QCLs)with central frequencies around 1900.0 cm~(-1),1600.0 cm~(-1),and 1103.4 cm~(-1)are used for NO,NO_2,and NH3detections,respectively,by timedivision multiplex.An open-path multi-pass cell of 60-m optical path length is applied to the instrument for high sensitivity and reducing the response time to less than 1 s.The prototype achieves a sub-ppb detection limit for all the three target gases with an average time of about 100 s.The instrument is installed in the Jiangsu environmental monitoring center to conduct performance tests on ambient air.Continuous 24-hour measurements show good agreement with the results of a reference instrument based on the chemiluminescence technique. |
| |
Keywords: | |
本文献已被 CNKI 等数据库收录! |
|