Abstract: | The performance of an integrated capillary electrophoresis system with a novel split-flow sample injection mechanism and special high sensitivity UV absorbance detector is described. Sample introduction into the capillary is accomplished with a standard HPLC-type microliter syringe. The injected sample is divided proportionally between the separation capillary and an adjustable splitvent. The volume of sample introduced into the capillary can be manipulated by varying the length or the i.d. of the splitvent tubing; or the volume of sample injected. Data are presented showing reproducibility of retention time, peak height, and peak area; minimum detectability; and operation at short UV wavelengths. |