An ion chromatographic separation method for the sequential determination of 90Sr, 241Am and Pu isotopes in a urine sample |
| |
Authors: | Baki B. Sadi Christopher Rinaldo Norman Spencer Chunsheng Li |
| |
Affiliation: | 1.Radiation Protection Bureau, Health Canada,Ottawa,Canada;2.Department of Biochemistry,University of Waterloo,Waterloo,Canada |
| |
Abstract: | ![]() A radiobioassay method has been developed for the sequential determination of 90Sr, 241Am and Pu isotopes in a urine sample. Unlike the existing methods using multiple extraction chromatographic cartridges, this work demonstrates an application of an automated ion chromatographic (IC) system for the separation of these radionuclides on a single IC column. The method meets the bioassay performance criteria for relative bias and relative precision as recommended by ANSI/HPS N13.30-2011. The detection limits for the radionuclides are found to be satisfactory for medical intervention in case of an accidental exposure scenario. Sample preparation time is less than 11 h. |
| |