首页 | 本学科首页   官方微博 | 高级检索  
     


Electrochemical biosensor based on self-assembled monolayers modified with gold nanoparticles for detection of HER-3
Authors:Mehmet Ç  etin Canbaz,Ç  iğdem Sayıklı ŞimşekMustafa Kemal Sezgintü  rk
Affiliation:Nam?k Kemal University, Faculty of Science and Arts, Chemistry Department, Biochemistry Division, Tekirda?, Turkey
Abstract:
We have developed a new immunological biosensor for ultrasensitive quantification of human epidermal growth factor receptor-3(HER-3). In order to construct the biosensor, the gold electrode surface was layered with, hexanedithiol, gold nanoparticles, and cysteamine, respectively. Anti-HER-3 antibody was covalently attached to cysteamine by glutaraldehyde and used as a bioreceptor in a biosensor system for the first time by this study. Surface characterization was obtained by means of electrochemical impedance spectroscopy and voltammetry. The proposed biosensor showed a good analytical performance for the detection of HER-3 ranging from 0.2 to 1.4 pg mL−1. Kramers–Kronig transform was performed on the experimental impedance data. Moreover, in an immunosensor system, the single frequency impedance technique was firstly used for characterization of interaction between HER-3 and anti-HER-3. Finally the presented biosensor was applied to artificial serum samples spiked with HER-3.
Keywords:Cancer biomarkers   HER-3   Anti-HER-3   Single frequency impedance   Biosensor
本文献已被 ScienceDirect 等数据库收录!