Electrochemical characterization of a primary electrolytic conductivity cell at CENAM |
| |
Authors: | A. Rodríguez-López A. Reyes-Del Valle J. M. Juárez-García M. Monroy-Mendoza M. J. Ávila-Salas J. L. Ortíz-Aparicio R. Antaño-López |
| |
Affiliation: | 1. Centro Nacional de Metrología, Km 4.5 Carretera a Los Cués, CP 76246, Municipio El Marques, Querétaro, Mexico 2. Centro de Investigación y Desarrollo Tecnológico en Electroquímica, S. C., Parque Tecnológico Querétaro—Sanfandila, CP 76703, Pedro Escobedo, Querétaro, Mexico
|
| |
Abstract: | Linear voltammetry and electrochemical impedance spectroscopy experiments were performed in order to obtain basic information about the electrochemical behavior of the primary cell of electrolytic conductivity at Centro Nacional de Metrología. Linear voltammetry shows that the amplitude of the sinusoidal perturbation must be smaller than 300 mV. Electrochemical impedance results indicate that optimal frequency interval extends from 5 to 0.6 kHz. Regarding sinusoidal signal amplitude, a lower bias with respect to electrolytic conductivity of an independent reference solution is obtained at 10 mV than at larger amplitudes; in addition, electrode integrity is improved. |
| |