首页 | 本学科首页   官方微博 | 高级检索  
     


Continuous chemiluminescence determination of formaldehyde in air based on Trautz-Schorigin reaction
Authors:Kamil Motyka,Pavel Miku&scaron  ka,Zbyněk Ve?e?a
Affiliation:a Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Veve?í 97, CZ-61142 Brno, Czech Republic
b Faculty of Science, Masaryk University, Kotlá?ská 2, CZ-61137 Brno, Czech Republic
Abstract:A new continuous method for the determination of formaldehyde in air is described. A cylindrical wet effluent diffusion denuder is used for the collection of formaldehyde from air into a thin film of absorption liquid (distilled-deionized water). Formaldehyde in the denuder concentrate is on-line detected employing a chemiluminescence flow method based on a reaction of formaldehyde and gallic acid with hydrogen peroxide in an alkaline solution. The collection efficiency of formaldehyde is quantitative at the air flow rate of 0.5 L min−1 (absorption liquid flow rate of 336 μL min−1). The limit of detection (S/N = 3) is 0.60 μg m−3 HCHO (0.49 ppb). The calibration graph is linear up to 300 μg m−3 HCHO (244 ppb). The relative standard deviations of chemiluminescence method for 1 × 10−6 and 5 × 10−6 M HCHO are 2.87% and 1.49%, respectively. Acetaldehyde interferes negligible, other compounds do not interfere. The method was employed for formaldehyde measurement in ambient air. The comparison measurement illustrates the good agreement of results obtained by proposed method with those obtained by reference fluorimetric method.
Keywords:Formaldehyde   Wet effluent diffusion denuder   Chemiluminescence   Air
本文献已被 ScienceDirect 等数据库收录!