首页 | 本学科首页   官方微博 | 高级检索  
     


Sensitive measurement of methylene blue active substances by attenuated total reflection spectrometry with a trimethylsilane-modified glass slab optical waveguide
Authors:Umemura Tomonari  Kasuya Yuichi  Odake Tamao  Tsunoda Kin-ichi
Affiliation:Department of Chemistry, Faculty of Engineering, Gunma University, Kiryu, Japan.
Abstract:
Attenuated total reflection spectrometry with a slab optical waveguide (SOWG) was explored for the simple, rapid and sensitive measurement of total anionic surfactants by the methylene blue active substance (MBAS) method. A fused-silica sheet used as a guiding layer was modified with trimethylsilane (TMS) to extract and concentrate the MBASs on the SOWG surface. Based on preliminary studies of the adsorption behavior and visible ATR spectrum of MB on the modified silica surface, a detection wavelength of 600 nm was chosen for the sensitive measurement of anionic surfactants. When the concentration of MB was set at 10 microM in the final measurement solution, the calibration curve for a typical anionic surfactant (sodium dodecylbenzenesulfonate) was linear up to 0.6 microM and the detection limit was 0.07 microM. The proposed method was applied to the determination of total anionic surfactants in river water.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号