Determination of arsenic(III), arsenic(V), antimony(III), antimony(V), selenium(IV) and selenium(VI) by extraction with ammonium pyrrolidinedithiocarbamate—methyl isobutyl ketone and electrothermal atomic absorption spectrometry |
| |
Authors: | Kunnath S Subramanian Jean C Meranger |
| |
Institution: | Environmental Health Centre, Health Protection Branch, Health and Welfare Canada, Tunney ''s Pasture. Ottawa, Ontario K1A0L2 Canada |
| |
Abstract: | The solution conditions and other parameters affecting the ammonium pyrrolidine-dithiocarbamate—methyl isobutyl ketone extraction system for graphite-furnace atomic absorption spectrometric determination of As(III), As(V), Sb(III), Sb(V), Se(IV) and Se(VI) were studied in detail. The solution conditions for the single or simultaneous extraction of As(III), Sb(III) and Se(IV) were not critical. Arsenic(V) and Se(VI) were not extracted over the entire range of pH and acidity studied. Antimony(V) was extracted only in the acidity range 0.3—1.0 M HCl. Simultaneous extraction of total arsenic and total antimony was possible after reduction of As(V) with thiosulphate. Interference studies are also reported. |
| |
Keywords: | |