Possibilities and limitations of the photoelectric aerosol sensor array applied to heavy metal aerosols |
| |
Authors: | Reinhard Niessner Beate Hemmerich Ulrich Panne |
| |
Affiliation: | (1) University of Dortmund, Inorganic and Analytical Chemistry, P. O. Box 500500, D-4600 Dortmund 50, Germany |
| |
Abstract: | Summary The technique of substance-selective aerosol charging by photoelectron emission was applied to several ultrafine heavy metal aerosols. To enhance the selectivity of the Photoelectric Aerosol Sensor (PAS), an array operating with four different wavelengths (185 nm, 214 nm, 229 nm, 254 nm) was used. The photoelectric activity at each wave-length and the linearity of the PAS signal was investigated for monodisperse and quasi-polydisperse systems. The aerosol mass concentration was determined by correlation with atomic absorption spectroscopy measurements of filter samples. Pattern recognition was applied to differentiate between four different aerosol systems and to identify common interferences.
Möglichkeiten und Grenzen des photoelektrischen Aerosolsensors für die Analyse von Schwermetallaerosolen Dedicated to Prof. Dr. G. Tölg on the occasion of his 60th birthday |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|