Development of a method for estimating an accurate equivalence point in nickel titration of cyanide ions |
| |
Authors: | Toshihiro Suzuki Masayasu Kurahashi |
| |
Affiliation: | National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), AIST Tsukuba Central 3-10, 1-1-1, Umezono, Tsukuba-shi, Ibaraki 305-8563, Japan |
| |
Abstract: | In a nickel titration of cyanide ions using murexide as indicator, an accurate equivalence point was determined by a non-linear least-squares curve-fitting for a titration curve. This method was developed to establish a standard solution for cyanide ions. In a curve-fitting procedure, a theoretical titration curve was calculated, assuming that nickel ion formed only a 1:4 Ni2+:CN− complex with cyanide ions and formed only a 1:1 complex with murexide. Results of the curve-fitting were reasonable at any pH and any indicator concentration studied. The combined standard uncertainty for a concentration of a 1000 mg kg−1 cyanide solution by this method was 0.079%. |
| |
Keywords: | Cyanide ion Complexometric titration Equivalence point Nickel ion Murexide Uncertainty |
本文献已被 ScienceDirect 等数据库收录! |
|