首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Response to high acidity and basicity at a platinum electrode in chronopotentiometry
Authors:Jianping Li  Tuzhi Peng
Institution:a Department of Chemistry, Institute of Microanalysis Systems, Xixi Campus, Zhejiang University, Hangzhou 310028, PR China
b Department of Applied Chemistry, Guilin Institute of Technology, Guilin 541004, PR China
Abstract:A simple and rapid method is developed to determine the high acidity and the basicity of solutions by chronopotentiometry with a platinum working electrode. The acidity range from 5.0 mol/l H+ to 1.0 mol/l OH can be measured by the adjustment of deposition potential and time. The response mechanism to acidity and basicity has been explored. The transition potential plateau in chronopotentiograms is caused from the oxidation of hydrogen adsorbed on electrode surface.
Keywords:Measurement of acidity  Strong acid  Strong base  Chronopotentiometry  Platinum electrode
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏