Serum estradiol measurement by solid-phase chemiluminescence immunoassay and direct radioimmunoassay |
| |
Authors: | Jozef de Boever Fortuna Kohen Marc Dhont Dirk Vandekerchkhove Georges van Maele |
| |
Affiliation: | Department of Obstetrics and Gynaecology, Academic Hospital, De Pintelaan 185, B-9000 Gent Belgium;Department of Medical Information, Academic Hospital, De Pintelaan 185, B-900 Gent Belgium |
| |
Abstract: | Estradiol017β is determined in serum extracts by solid-phase chemiluminescence immunoassay. The results are compared with those obtained from unextracted serum in routine conditions with a commercial radioimmunoassay (r.i.a.) kit. For the chemiluminescence procedure, a purified monoclonal antibody to estradiol-6-carboxymethyloxime/bovine serum albumin and the homologous chemiluminescent marker conjugate estradiol-6-carboxymethyloxime aminobutylethylisoluminol are used. Bound and free ligand are separated by washing and simple centrifugation. Results obtained by the chemiluminescence assay (y) and by r.i.a. (x) on 170 serum specimens from women during ovulation induction showed good correlation (y = 1.01x ? 16 with r = 0.95). The methods are similar in selectivity, detection limit (ca. 10 ng l?1) and precision (interassay relative standard deviation, 8–13%). |
| |