首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of diclofenac sodium in eagle's minimum essential medium with Earle's balanced salt solution
Authors:Lopes Patrícia S  Kaneko Telma M  Takano Carolina Y  Lacerda Aurea C L  Latorre Leandro R  Kato Massuo J
Affiliation:Faculty of Pharmaceutical Sciences Oswaldo Cruz, R. Brigadeiro Galv?o 540, 01151-000, S?o Paulo, SP, Brazil.
Abstract:A validated method was developed for determination of diclofenac sodium, considered a model hydrophilic drug for in vitro permeation studies, in Eagle's Minimum Essential Medium (MEM) with Earle's balanced salt solution. Liquid chromatography was used to determine diclofenac sodium. This method was developed with a reversed-phase column Supercosil LC 18, DB 25 cm x 4.5 mm; the mobile phase was methanol with 3% (v/v) acetic acid-Milli-Q water (74 + 26), and detection was at 283 nm. The detection and quantitation limits were 2.41 x 10(-8) and 3.31 x 10(-5) microg/microL, respectively. The accuracy within-day (n = 3) and day-to-day (n = 7) was 98.83%; the mean variation coefficient for inter- (n = 7) and intraday precision (n = 3) was 12.20%, thus, not exceeding 15%. This method can be used as an analytical procedure for the determination of diclofenac sodium in MEM for in vitro permeation studies.
Keywords:
本文献已被 PubMed 等数据库收录!