首页 | 本学科首页   官方微博 | 高级检索  
     


An electronic aid for line selection for scanning inductively coupled plasma atomic emission spectrometers
Authors:Douglas P. Webb  Eric D. Salin  
Affiliation:

Department of Chemistry, McGill University, 801 Sherbrooke St. W., Montreal, P.Q., Canada H3A 2K6

Abstract:
This article is an electronic publication in Spectrochimica Acta Electronica (SAE), the electronic section of Spectrochimica Acta Part B (SAB). Accompanying this hard copy text is a disk containing the program, source flies, data files and a brief manual along with a few programming notes. The main article discusses the purpose of this work, and the Appendix provides brief instruction on the use of the program, manipulation of the data, and source code. An automatic file selection system for the control of a rapid scanning spectrometer performing atomic emission spectrometry has been developed using the PROLOG language. The system can be “taught” the elemental composition of a given sample type. It uses this information (when available) to generate a linear search strategy which minimizes the number of lines that need to be measured. The rules used by the system can be observed during system operation to allow a trace of the logic. Reading of the initial data base takes no more than 6 s and then decisions are made in less than 1 s per line.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号