Simple simulation method for investigating the performance of a SQUID gradiometer corresponding to a baseline length |
| |
Authors: | Chan Seok Kang Kwon Kyu Yu Kiwoong Kim Hyukchan Kwon Jin-Mok Kim Yong-Ho Lee |
| |
Affiliation: | Brain and Cognition Measurement Lab, Korea Research Institute of Standards and Science, 267 Gajeong-ro, Yuseong-gu, Daejeon 305-340, Republic of Korea |
| |
Abstract: | Since the Superconducting Quantum Interference Device (SQUID) is the most sensitive magnetic field detector, it is widely used for measuring very weak biomagnetic signals, such as a Magnetocardiography (MCG). However, because the SQUID is easily affected by environmental magnetic noises, it is often used as a form of gradiometer suppressing spatially homogeneous field noises. The characteristic of the SQUID gradiometer mainly depends on the distance between the connected coils, that is, a baseline. Accordingly, in order to fabricate a gradiometer, the investigation of a proper baseline for the MCG signal is necessary. For the investigation, however, various gradiometers having different baselines have to be fabricated and tested, which required too much cost and effort. In this study, we suggest a simple simulation method to investigate the performance of a SQUID gradiometer depending on the baseline. In addition, in order to verify the reliability of our simulation, we fabricated wire wound axial first-order gradiometers (50 mm, 70 mm, and 100 mm baseline) and a second-order gradiometer (50 mm baseline) and compared the characteristics of the measured gradiometer signal with the simulated signal. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|