首页 | 本学科首页   官方微博 | 高级检索  
     


Optimization of detection device geometry for NIR spectroscopy using a three-layered model of stone fruit
Authors:Zuojun Tan  Jing Xie  Jianjun Chen  Chizhu Ding
Affiliation:1.Huazhong Agricultural University,College of Science,Wuhan,China
Abstract:
The influences of detection device geometry and fiber optic parameters on near infrared spectroscopy measurements were assessed using stone fruit models based on Monte Carlo simulation. The stone fruit was modeled as concentric spherical layered tissues including the skin, the flesh and the core. The choices of the detection angle, the diameter of the detection fiber, the numerical aperture, and the height of the probe were discussed. Receiving diffuse reflectance signals at detection angles in the range of 35°–50° and normalizing the detection signals by the collection area and the solid acceptance angle prior to use are suggested. Fiber probes with diameters D = 0.06 cm or 0.1 cm, NA = 0.20 or 0.30, and height h ≤ 0.8 cm are preferred. The probe deflection angle should be limited to within ±5° to guarantee measurement accuracy.
Keywords:
本文献已被 SpringerLink 等数据库收录!