Abstract: | The choice of the reference, either as internal or external is not straightforward in quantitative NMR. In this context ERETIC™ methodology appears as an universal referencing technique. An electronic signal, generated by the NMR spectrometer during the acquisition time, operates as a virtual working standard. The processes for ensuring a traceability to primary standards is illustrated on the official method devoted to (D/H)i ratios measurement on ethanol, using quantitative 2H-NMR. The ERETIC approach is shown to be equivalent to its official homologue, in terms of accuracy and precision. Finally, its performance could be beneficial to other analytes, matrices and nuclei. |
| |