Estimation of uranium in bioassay samples of occupational workers by laser fluorimetry
Authors:A Suja  S P Prabhu  P D Sawant  A K Tiwari  R Sharma  P K Sarkar
Institution:1. Health Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai, 400 085, India
2. Chemical Technology Division, Bhabha Atomic Research Centre, Trombay, Mumbai, 400 085, India
3. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Trombay, Mumbai, 400 085, India
Abstract:A newly established uranium processing facility has been commissioned at BARC, Trombay. Monitoring of occupational workers is essential to assess intake of uranium in this facility. A group of 21 workers was selected for bioassay monitoring to assess the existing urinary excretion levels of uranium before the commencement of actual work. Bioassay samples collected from these workers were analyzed by ion-exchange technique followed by laser fluorimetry. Standard addition method was followed for estimation of uranium concentration in the samples. The minimum detectable activity by this technique is about 0.2 ng. The range of uranium observed in these samples varies from 19 to 132 ng/L. Few of these samples were also analyzed by fission track analysis technique and the results were found to be comparable to those obtained by laser fluorimetry. The urinary excretion rate observed for the individual can be regarded as a ‘personal baseline’ and will be treated as the existing level of uranium in urine for these workers at the facility.